IPS pressure ranges |
- 25, 70, 200, 350 and 700 mbar gauge
- 1, 2, 3.5, 7, 10, 20, 35, 70, 100, 135, 172, 210 bar gauge
- 350, 420, 700 and 1000 bar absolute
- 0.35, 1, 3, 5, 10, 15, 30, 50, 100, 150, 300, 500, 1000, 1500, 2000, 2500, 3000 psi gauge
- 5000, 6000, 10000 and 14500 psi absolute
- 2.5, 7, 20, 35, 70, 100, 200, 350, 700 kPa gauge
- 1, 2, 3.5, 7, 10, 13.5, 17.2, 21 MPa gauge
- 35, 42, 70 and 100 MPa absolute
|
IRS pressure ranges |
- 1, 2.5, 7, 10, 20, 35 and 70, 100, 135, 172, 210 bar pseudo gauge
- 2, 3.5, 8, 11, 21, 36, 71, 101, 136, 173 and 211 bar absolute
- 15, 36, 101, 145, 290, 507, 1015, 1450, 1958, 2494, 3046 psi pseudo gauge
- 29, 44, 73, 116, 160, 305, 522, 1030, 1465, 1973, 2509, 3060 psi absolute
- 0.1, 0.25, 0.7, 1.2, 3.5, 7, 10, 13.5, 17.2 21 MPa pseudo gauge
- 0.2, 0.3, 0.5, 0.8, 1.1, 2.1, 3.6, 7.1, 10.1, 13.6, 17.3, 21.1 MPa absolute
|
IRS-B barometric reference ranges: |
750-1150 mbar absolute, 10.9-16.7 psi absolute, 75-115 kPa absolute |
Display panel |
¼ VGA wide format 4.3 inch colour graphics LCD c/w integral touch screen |
Comms update rate |
8 times per second |
IPS0 standard precision |
- 25 mbar: 0.20% Rdg + 0.20% FS
- 70 mbar: 0.10% Rdg + 0.10% FS
- 200 mbar: 0.04% Rdg + 0.04% FS
- Other ranges: 0.02% Rdg + 0.02% FS
Includes linearity, hysteresis, repeatability and temperature effects,
assumes steady state temperature and regular zeroing (for absolute
ranges an IRS-B needs to be installed)
|
IPS1 high precision |
- 25 mbar: 0.10% Rdg + 0.10% FS
- 70 mbar: 0.05% Rdg + 0.05% FS
- 200 mbar: 0.02% Rdg + 0.02% FS
- Other ranges: 0.01% Rdg + 0.01% FS
Includes linearity, hysteresis, repeatability and temperature effects,
assumes steady state temperature and regular zeroing (for absolute
ranges an IRS-B needs to be installed)
|
IPS2 high precision |
- 25 mbar: 0.05% Rdg + 0.05% FS
- 70 mbar: 0.025% Rdg + 0.025% FS
- 200 mbar: 0.01% Rdg + 0.01% FS
- Other ranges: 0.005% Rdg + 0.005% FS
Includes linearity, hysteresis, repeatability and temperature effects,
assumes steady state temperature and regular zeroing (for absolute
ranges an IRS-B needs to be installed)
|
Power supply |
90 VAC to 130 VAC @ 47 to 63 Hz and 180 VAC to 260 VAC @ 47 to 63 Hz. 15 VA |
Communication |
RS232, USB and IEEE-488, SCPI, DPI141, DPI142 and DPI150 emulation.
LabVIEW drivers Ethernet (VXI-II and Sockets using SCPI) |